Validation of a point of care troponin assay in real life emergency department conditions. Loten, C., Attia, J., Hullick, C., Marley, J., & McElduff, P. Emergency Medicine Australasia, 21(4):286–292, August, 2009.
Validation of a point of care troponin assay in real life emergency department conditions [link]Paper  doi  abstract   bibtex   
Objective:  To validate the accuracy of a Point of care (POC) troponin device (Abbott i-Stat) in real life ED conditions. Methods:  Design: A three-way comparison between troponin I results obtained by experienced POC operators, inexperienced ED staff and central laboratory criterion standard. Participants: Convenience sample of 332 patients presenting to the ED with possible coronary syndromes. Results:  Spearman correlation coefficient for experienced versus laboratory was 0.83 (95% CI 0.78–0.87), occasional users versus laboratory was 0.76 (95% CI 0.71–0.81), and experienced versus occasional users on POC was 0.82 (95% CI 0.76–0.87). Using local troponin cut-off of 0.1 ng/mL, kappa coefficient was 0.94 for occasional users versus laboratory, 0.91 for experienced versus laboratory and 0.94 for experienced versus occasional users. Bland–Altman plots showed good agreement across the range of measured values. The sensitivity of i-Stat (vs laboratory as criterion standard) was 92.2% (95% CI 83.8%–97.0%) with the local cut-off but only 70.1% (95% CI 60.5%–78.6%) using the lowest cut-off associated with acceptable reproducibility (10% coefficient of variation). Conclusions:  The i-Stat POC device produces similar results in the hands of experienced and occasional operators in ED. There is good agreement between the POC and laboratory at levels used to diagnose infarction by older, more specific criteria. When compared using new lower cut-offs, the i-Stat had poor sensitivity.
@article{loten_validation_2009,
	title = {Validation of a point of care troponin assay in real life emergency department conditions},
	volume = {21},
	issn = {1742-6723},
	url = {http://onlinelibrary.wiley.com.ezproxy.otago.ac.nz/doi/10.1111/j.1742-6723.2009.01198.x/abstract},
	doi = {10.1111/j.1742-6723.2009.01198.x},
	abstract = {Objective:  To validate the accuracy of a Point of care (POC) troponin device (Abbott i-Stat) in real life ED conditions. Methods:  Design: A three-way comparison between troponin I results obtained by experienced POC operators, inexperienced ED staff and central laboratory criterion standard. Participants: Convenience sample of 332 patients presenting to the ED with possible coronary syndromes. Results:  Spearman correlation coefficient for experienced versus laboratory was 0.83 (95\% CI 0.78–0.87), occasional users versus laboratory was 0.76 (95\% CI 0.71–0.81), and experienced versus occasional users on POC was 0.82 (95\% CI 0.76–0.87). Using local troponin cut-off of 0.1 ng/mL, kappa coefficient was 0.94 for occasional users versus laboratory, 0.91 for experienced versus laboratory and 0.94 for experienced versus occasional users. Bland–Altman plots showed good agreement across the range of measured values. The sensitivity of i-Stat (vs laboratory as criterion standard) was 92.2\% (95\% CI 83.8\%–97.0\%) with the local cut-off but only 70.1\% (95\% CI 60.5\%–78.6\%) using the lowest cut-off associated with acceptable reproducibility (10\% coefficient of variation). Conclusions:  The i-Stat POC device produces similar results in the hands of experienced and occasional operators in ED. There is good agreement between the POC and laboratory at levels used to diagnose infarction by older, more specific criteria. When compared using new lower cut-offs, the i-Stat had poor sensitivity.},
	language = {en},
	number = {4},
	urldate = {2018-03-17TZ},
	journal = {Emergency Medicine Australasia},
	author = {Loten, Conrad and Attia, John and Hullick, Carolyn and Marley, John and McElduff, Patrick},
	month = aug,
	year = {2009},
	keywords = {\_tablet, emergency medicine, point of care system, troponin},
	pages = {286--292}
}

Downloads: 0