Comparative evaluation of methodologies for T-wave alternans mapping in electrograms. Orini, M., Hanson, B., Monasterio, V., Martínez, J., P., Hayward, M., Taggart, P., & Lambiase, P. IEEE Transactions on Biomedical Engineering, 61(2):308-316, 2014. Paper abstract bibtex Electrograms (EGM) recorded from the surface of the myocardium are becoming more and more accessible. T-wave alternans (TWA) is associated with increased vulnerability to ventricular tachycardia/fibrillation and it occurs before the onset of ventricular arrhythmias. Thus, accurate methodologies for time-varying alternans estimation/detection in EGM are needed. In this paper, we perform a simulation study based on epicardial EGM recorded in vivo in humans to compare the accuracy of four methodologies: the spectral method (SM), modified moving average method (MMA), laplacian likelihood ratio method (LLR) and a novel method based on time-frequency distributions (TFD). A variety of effects are considered, which include the presence of wide band noise, respiration and impulse artifacts. We found that (a) EGM-TWA can be detected accurately when the standard deviation of wide band noise is equal or smaller than 10 times the magnitude of EGM-TWA. (b) Respiration can be critical for EGM-TWA analysis, even at typical respiratory rates. (c) Impulse noise strongly reduces the accuracy of all methods, except LLR. (d) If depolarization time is used as a fiducial point, the localization of the T-wave is not critical for the accuracy of EGMTWA detection. (e) According to this study, all methodologies provided accurate EGM-TWA detection/quantification in ideal conditions, while LLR was the most robust, providing better detection-rates in noisy conditions. Application on epicardial mapping of the in-vivo human heart shows that EGM-TWA has heterogeneous spatio-temporal distribution.
@article{
title = {Comparative evaluation of methodologies for T-wave alternans mapping in electrograms},
type = {article},
year = {2014},
identifiers = {[object Object]},
keywords = {Electrograms (EGM),T-wave alternans,intracardiac ECG,repolarization,ventricular arrhythmia},
pages = {308-316},
volume = {61},
id = {e3469e8a-d744-3f07-896f-2c11459100ea},
created = {2015-04-30T11:14:47.000Z},
file_attached = {true},
profile_id = {5e73f9af-1f10-3a70-a285-d3fbb699efe2},
group_id = {e134fb94-8118-3ebe-b82d-bfbf0a8ae633},
last_modified = {2015-04-30T12:49:41.000Z},
read = {false},
starred = {false},
authored = {false},
confirmed = {true},
hidden = {false},
abstract = {Electrograms (EGM) recorded from the surface of the myocardium are becoming more and more accessible. T-wave alternans (TWA) is associated with increased vulnerability to ventricular tachycardia/fibrillation and it occurs before the onset of ventricular arrhythmias. Thus, accurate methodologies for time-varying alternans estimation/detection in EGM are needed. In this paper, we perform a simulation study based on epicardial EGM recorded in vivo in humans to compare the accuracy of four methodologies: the spectral method (SM), modified moving average method (MMA), laplacian likelihood ratio method (LLR) and a novel method based on time-frequency distributions (TFD). A variety of effects are considered, which include the presence of wide band noise, respiration and impulse artifacts. We found that (a) EGM-TWA can be detected accurately when the standard deviation of wide band noise is equal or smaller than 10 times the magnitude of EGM-TWA. (b) Respiration can be critical for EGM-TWA analysis, even at typical respiratory rates. (c) Impulse noise strongly reduces the accuracy of all methods, except LLR. (d) If depolarization time is used as a fiducial point, the localization of the T-wave is not critical for the accuracy of EGMTWA detection. (e) According to this study, all methodologies provided accurate EGM-TWA detection/quantification in ideal conditions, while LLR was the most robust, providing better detection-rates in noisy conditions. Application on epicardial mapping of the in-vivo human heart shows that EGM-TWA has heterogeneous spatio-temporal distribution.},
bibtype = {article},
author = {Orini, Michele and Hanson, Ben and Monasterio, Violeta and Martínez, Juan Pablo and Hayward, Martin and Taggart, Peter and Lambiase, Pier},
journal = {IEEE Transactions on Biomedical Engineering},
number = {2}
}
Downloads: 0
{"_id":"QT3zXhLcpt3uCdKmv","bibbaseid":"orini-hanson-monasterio-martnez-hayward-taggart-lambiase-comparativeevaluationofmethodologiesfortwavealternansmappinginelectrograms-2014","downloads":0,"creationDate":"2016-05-08T13:37:40.043Z","title":"Comparative evaluation of methodologies for T-wave alternans mapping in electrograms","author_short":["Orini, M.","Hanson, B.","Monasterio, V.","Martínez, J., P.","Hayward, M.","Taggart, P.","Lambiase, P."],"year":2014,"bibtype":"article","biburl":null,"bibdata":{"title":"Comparative evaluation of methodologies for T-wave alternans mapping in electrograms","type":"article","year":"2014","identifiers":"[object Object]","keywords":"Electrograms (EGM),T-wave alternans,intracardiac ECG,repolarization,ventricular arrhythmia","pages":"308-316","volume":"61","id":"e3469e8a-d744-3f07-896f-2c11459100ea","created":"2015-04-30T11:14:47.000Z","file_attached":"true","profile_id":"5e73f9af-1f10-3a70-a285-d3fbb699efe2","group_id":"e134fb94-8118-3ebe-b82d-bfbf0a8ae633","last_modified":"2015-04-30T12:49:41.000Z","read":false,"starred":false,"authored":false,"confirmed":"true","hidden":false,"abstract":"Electrograms (EGM) recorded from the surface of the myocardium are becoming more and more accessible. T-wave alternans (TWA) is associated with increased vulnerability to ventricular tachycardia/fibrillation and it occurs before the onset of ventricular arrhythmias. Thus, accurate methodologies for time-varying alternans estimation/detection in EGM are needed. In this paper, we perform a simulation study based on epicardial EGM recorded in vivo in humans to compare the accuracy of four methodologies: the spectral method (SM), modified moving average method (MMA), laplacian likelihood ratio method (LLR) and a novel method based on time-frequency distributions (TFD). A variety of effects are considered, which include the presence of wide band noise, respiration and impulse artifacts. We found that (a) EGM-TWA can be detected accurately when the standard deviation of wide band noise is equal or smaller than 10 times the magnitude of EGM-TWA. (b) Respiration can be critical for EGM-TWA analysis, even at typical respiratory rates. (c) Impulse noise strongly reduces the accuracy of all methods, except LLR. (d) If depolarization time is used as a fiducial point, the localization of the T-wave is not critical for the accuracy of EGMTWA detection. (e) According to this study, all methodologies provided accurate EGM-TWA detection/quantification in ideal conditions, while LLR was the most robust, providing better detection-rates in noisy conditions. Application on epicardial mapping of the in-vivo human heart shows that EGM-TWA has heterogeneous spatio-temporal distribution.","bibtype":"article","author":"Orini, Michele and Hanson, Ben and Monasterio, Violeta and Martínez, Juan Pablo and Hayward, Martin and Taggart, Peter and Lambiase, Pier","journal":"IEEE Transactions on Biomedical Engineering","number":"2","bibtex":"@article{\n title = {Comparative evaluation of methodologies for T-wave alternans mapping in electrograms},\n type = {article},\n year = {2014},\n identifiers = {[object Object]},\n keywords = {Electrograms (EGM),T-wave alternans,intracardiac ECG,repolarization,ventricular arrhythmia},\n pages = {308-316},\n volume = {61},\n id = {e3469e8a-d744-3f07-896f-2c11459100ea},\n created = {2015-04-30T11:14:47.000Z},\n file_attached = {true},\n profile_id = {5e73f9af-1f10-3a70-a285-d3fbb699efe2},\n group_id = {e134fb94-8118-3ebe-b82d-bfbf0a8ae633},\n last_modified = {2015-04-30T12:49:41.000Z},\n read = {false},\n starred = {false},\n authored = {false},\n confirmed = {true},\n hidden = {false},\n abstract = {Electrograms (EGM) recorded from the surface of the myocardium are becoming more and more accessible. T-wave alternans (TWA) is associated with increased vulnerability to ventricular tachycardia/fibrillation and it occurs before the onset of ventricular arrhythmias. Thus, accurate methodologies for time-varying alternans estimation/detection in EGM are needed. In this paper, we perform a simulation study based on epicardial EGM recorded in vivo in humans to compare the accuracy of four methodologies: the spectral method (SM), modified moving average method (MMA), laplacian likelihood ratio method (LLR) and a novel method based on time-frequency distributions (TFD). A variety of effects are considered, which include the presence of wide band noise, respiration and impulse artifacts. We found that (a) EGM-TWA can be detected accurately when the standard deviation of wide band noise is equal or smaller than 10 times the magnitude of EGM-TWA. (b) Respiration can be critical for EGM-TWA analysis, even at typical respiratory rates. (c) Impulse noise strongly reduces the accuracy of all methods, except LLR. (d) If depolarization time is used as a fiducial point, the localization of the T-wave is not critical for the accuracy of EGMTWA detection. (e) According to this study, all methodologies provided accurate EGM-TWA detection/quantification in ideal conditions, while LLR was the most robust, providing better detection-rates in noisy conditions. Application on epicardial mapping of the in-vivo human heart shows that EGM-TWA has heterogeneous spatio-temporal distribution.},\n bibtype = {article},\n author = {Orini, Michele and Hanson, Ben and Monasterio, Violeta and Martínez, Juan Pablo and Hayward, Martin and Taggart, Peter and Lambiase, Pier},\n journal = {IEEE Transactions on Biomedical Engineering},\n number = {2}\n}","author_short":["Orini, M.","Hanson, B.","Monasterio, V.","Martínez, J., P.","Hayward, M.","Taggart, P.","Lambiase, P."],"urls":{"Paper":"http://bibbase.org/service/mendeley/5e73f9af-1f10-3a70-a285-d3fbb699efe2/file/7ec6a1bd-d994-6b65-53c1-dc2ec6ca0fc2/2014-Comparative_evaluation_of_methodologies_for_T-wave_alternans_mapping_in_electrograms.pdf.pdf"},"bibbaseid":"orini-hanson-monasterio-martnez-hayward-taggart-lambiase-comparativeevaluationofmethodologiesfortwavealternansmappinginelectrograms-2014","role":"author","keyword":["Electrograms (EGM)","T-wave alternans","intracardiac ECG","repolarization","ventricular arrhythmia"],"downloads":0},"search_terms":["comparative","evaluation","methodologies","wave","alternans","mapping","electrograms","orini","hanson","monasterio","martínez","hayward","taggart","lambiase"],"keywords":["electrograms (egm)","t-wave alternans","intracardiac ecg","repolarization","ventricular arrhythmia"],"authorIDs":[]}