Model fitting of volumetric capnograms improves calculations of airway dead space and slope of phase III. Tusman, G., Scandurra, A., Böhm, S., H., Suarez-Sipmann, F., & Clara, F. Journal of clinical monitoring and computing, 23(4):197-206, 8, 2009.
Paper abstract bibtex BACKGROUND: This study assessed the performance of a Functional Approximation based on a Levenberg-Marquardt Algorithm (FA-LMA) to calculate airway dead space (VD(aw)) and the slope of phase III (S(III)) from capnograms. METHODS: We performed mathematical simulations to test the effect of noises on the calculation of VD(aw) and S(III). Data from ten mechanically ventilated patients at 0, 5 and 10 cmH(2)O of PEEP were also studied. FA-LMA was compared with the traditional Fowler's method (FM). RESULTS: Simulations showed that: (1) The FM determined VD(aw) with accuracy only if the capnogram approximated a symmetrical curve (S(III) = 0). When capnograms became asymmetrical (S(III) > 0), the FM underestimated VD(aw) (-3.1% to -0.9%). (2) When adding noises on 800 capnograms, VD(aw) was underestimated whenever the FM was used thereby creating a bias between -5.54 and -1.28 ml at standard deviations (SD) of 0.1-1.8 ml (P < 0.0001). FA-LMA calculations of VD(aw) were close to the simulated values with the bias ranging from -0.21 to 0.16 ml at SD from 0.1 to 0.4 ml. The FM overestimated S(III) and showed more bias (0.0041-0.0078 mmHg/ml, P < 0.0001) than the FA-LMA (0.0002-0.0030 mmHg/ml). When calculating VD(aw) from patients, variability was less with the FA-LMA leading to mean variation coefficients of 0.0102, 0.0111 and 0.0123 compared to the FM (0.0243, 0.0247 and 0.0262, P < 0.001) for 0, 5 and 10 cmH(2)O of PEEP, respectively. The FA-LMA also showed less variability in S(III) with mean variation coefficients of 0.0739, 0.0662 and 0.0730 compared to the FM (0.1379, 0.1208 and 0.1246, P < 0.001) for 0, 5 and 10 cmH(2)O of PEEP, respectively. CONCLUSIONS: The Functional Approximation based on a Levenberg-Marquardt Algorithm showed less bias and dispersion compared to the traditional Fowler's method when calculating VD(aw) and S(III).
@article{
id = {f805ebb2-9f21-3395-9fd6-f73114aaa1a7},
title = {Model fitting of volumetric capnograms improves calculations of airway dead space and slope of phase III.},
type = {article},
year = {2009},
identifiers = {[object Object]},
keywords = {abstract,background,co 2,dead space,from the 1 department,hospital privado de,lung recruit-,ment,monitoring,of anesthesiology,peep,phase iii slope,volumetric capnography},
created = {2009-08-17T18:00:51.000Z},
pages = {197-206},
volume = {23},
websites = {http://www.ncbi.nlm.nih.gov/pubmed/19517259},
month = {8},
file_attached = {true},
profile_id = {fe7067eb-58b8-34c6-b8cd-6717fdf7605c},
group_id = {f87494e7-529d-3e68-b1f1-0e6d57d138db},
last_modified = {2014-07-19T19:15:34.000Z},
read = {true},
starred = {true},
authored = {false},
confirmed = {true},
hidden = {false},
citation_key = {Tusman2009},
client_data = {"desktop_id":"4e6295bb-e114-4798-bba2-2078081119f5"},
abstract = {BACKGROUND: This study assessed the performance of a Functional Approximation based on a Levenberg-Marquardt Algorithm (FA-LMA) to calculate airway dead space (VD(aw)) and the slope of phase III (S(III)) from capnograms. METHODS: We performed mathematical simulations to test the effect of noises on the calculation of VD(aw) and S(III). Data from ten mechanically ventilated patients at 0, 5 and 10 cmH(2)O of PEEP were also studied. FA-LMA was compared with the traditional Fowler's method (FM). RESULTS: Simulations showed that: (1) The FM determined VD(aw) with accuracy only if the capnogram approximated a symmetrical curve (S(III) = 0). When capnograms became asymmetrical (S(III) > 0), the FM underestimated VD(aw) (-3.1% to -0.9%). (2) When adding noises on 800 capnograms, VD(aw) was underestimated whenever the FM was used thereby creating a bias between -5.54 and -1.28 ml at standard deviations (SD) of 0.1-1.8 ml (P < 0.0001). FA-LMA calculations of VD(aw) were close to the simulated values with the bias ranging from -0.21 to 0.16 ml at SD from 0.1 to 0.4 ml. The FM overestimated S(III) and showed more bias (0.0041-0.0078 mmHg/ml, P < 0.0001) than the FA-LMA (0.0002-0.0030 mmHg/ml). When calculating VD(aw) from patients, variability was less with the FA-LMA leading to mean variation coefficients of 0.0102, 0.0111 and 0.0123 compared to the FM (0.0243, 0.0247 and 0.0262, P < 0.001) for 0, 5 and 10 cmH(2)O of PEEP, respectively. The FA-LMA also showed less variability in S(III) with mean variation coefficients of 0.0739, 0.0662 and 0.0730 compared to the FM (0.1379, 0.1208 and 0.1246, P < 0.001) for 0, 5 and 10 cmH(2)O of PEEP, respectively. CONCLUSIONS: The Functional Approximation based on a Levenberg-Marquardt Algorithm showed less bias and dispersion compared to the traditional Fowler's method when calculating VD(aw) and S(III).},
bibtype = {article},
author = {Tusman, Gerardo and Scandurra, Adriana and Böhm, Stephan H and Suarez-Sipmann, Fernando and Clara, Fernando},
journal = {Journal of clinical monitoring and computing},
number = {4}
}
Downloads: 0
{"_id":"cBQ5A3Y9qfWtGyu3b","authorIDs":[],"author_short":["Tusman, G.","Scandurra, A.","Böhm, S., H.","Suarez-Sipmann, F.","Clara, F."],"bibbaseid":"tusman-scandurra-bhm-suarezsipmann-clara-modelfittingofvolumetriccapnogramsimprovescalculationsofairwaydeadspaceandslopeofphaseiii-2009","bibdata":{"id":"f805ebb2-9f21-3395-9fd6-f73114aaa1a7","title":"Model fitting of volumetric capnograms improves calculations of airway dead space and slope of phase III.","type":"article","year":"2009","identifiers":"[object Object]","keywords":"abstract,background,co 2,dead space,from the 1 department,hospital privado de,lung recruit-,ment,monitoring,of anesthesiology,peep,phase iii slope,volumetric capnography","created":"2009-08-17T18:00:51.000Z","pages":"197-206","volume":"23","websites":"http://www.ncbi.nlm.nih.gov/pubmed/19517259","month":"8","file_attached":"true","profile_id":"fe7067eb-58b8-34c6-b8cd-6717fdf7605c","group_id":"f87494e7-529d-3e68-b1f1-0e6d57d138db","last_modified":"2014-07-19T19:15:34.000Z","read":"true","starred":"true","authored":"false","confirmed":"true","hidden":"false","citation_key":"Tusman2009","client_data":"\"desktop_id\":\"4e6295bb-e114-4798-bba2-2078081119f5\"","abstract":"BACKGROUND: This study assessed the performance of a Functional Approximation based on a Levenberg-Marquardt Algorithm (FA-LMA) to calculate airway dead space (VD(aw)) and the slope of phase III (S(III)) from capnograms. METHODS: We performed mathematical simulations to test the effect of noises on the calculation of VD(aw) and S(III). Data from ten mechanically ventilated patients at 0, 5 and 10 cmH(2)O of PEEP were also studied. FA-LMA was compared with the traditional Fowler's method (FM). RESULTS: Simulations showed that: (1) The FM determined VD(aw) with accuracy only if the capnogram approximated a symmetrical curve (S(III) = 0). When capnograms became asymmetrical (S(III) > 0), the FM underestimated VD(aw) (-3.1% to -0.9%). (2) When adding noises on 800 capnograms, VD(aw) was underestimated whenever the FM was used thereby creating a bias between -5.54 and -1.28 ml at standard deviations (SD) of 0.1-1.8 ml (P < 0.0001). FA-LMA calculations of VD(aw) were close to the simulated values with the bias ranging from -0.21 to 0.16 ml at SD from 0.1 to 0.4 ml. The FM overestimated S(III) and showed more bias (0.0041-0.0078 mmHg/ml, P < 0.0001) than the FA-LMA (0.0002-0.0030 mmHg/ml). When calculating VD(aw) from patients, variability was less with the FA-LMA leading to mean variation coefficients of 0.0102, 0.0111 and 0.0123 compared to the FM (0.0243, 0.0247 and 0.0262, P < 0.001) for 0, 5 and 10 cmH(2)O of PEEP, respectively. The FA-LMA also showed less variability in S(III) with mean variation coefficients of 0.0739, 0.0662 and 0.0730 compared to the FM (0.1379, 0.1208 and 0.1246, P < 0.001) for 0, 5 and 10 cmH(2)O of PEEP, respectively. CONCLUSIONS: The Functional Approximation based on a Levenberg-Marquardt Algorithm showed less bias and dispersion compared to the traditional Fowler's method when calculating VD(aw) and S(III).","bibtype":"article","author":"Tusman, Gerardo and Scandurra, Adriana and Böhm, Stephan H and Suarez-Sipmann, Fernando and Clara, Fernando","journal":"Journal of clinical monitoring and computing","number":"4","bibtex":"@article{\n id = {f805ebb2-9f21-3395-9fd6-f73114aaa1a7},\n title = {Model fitting of volumetric capnograms improves calculations of airway dead space and slope of phase III.},\n type = {article},\n year = {2009},\n identifiers = {[object Object]},\n keywords = {abstract,background,co 2,dead space,from the 1 department,hospital privado de,lung recruit-,ment,monitoring,of anesthesiology,peep,phase iii slope,volumetric capnography},\n created = {2009-08-17T18:00:51.000Z},\n pages = {197-206},\n volume = {23},\n websites = {http://www.ncbi.nlm.nih.gov/pubmed/19517259},\n month = {8},\n file_attached = {true},\n profile_id = {fe7067eb-58b8-34c6-b8cd-6717fdf7605c},\n group_id = {f87494e7-529d-3e68-b1f1-0e6d57d138db},\n last_modified = {2014-07-19T19:15:34.000Z},\n read = {true},\n starred = {true},\n authored = {false},\n confirmed = {true},\n hidden = {false},\n citation_key = {Tusman2009},\n client_data = {\"desktop_id\":\"4e6295bb-e114-4798-bba2-2078081119f5\"},\n abstract = {BACKGROUND: This study assessed the performance of a Functional Approximation based on a Levenberg-Marquardt Algorithm (FA-LMA) to calculate airway dead space (VD(aw)) and the slope of phase III (S(III)) from capnograms. METHODS: We performed mathematical simulations to test the effect of noises on the calculation of VD(aw) and S(III). Data from ten mechanically ventilated patients at 0, 5 and 10 cmH(2)O of PEEP were also studied. FA-LMA was compared with the traditional Fowler's method (FM). RESULTS: Simulations showed that: (1) The FM determined VD(aw) with accuracy only if the capnogram approximated a symmetrical curve (S(III) = 0). When capnograms became asymmetrical (S(III) > 0), the FM underestimated VD(aw) (-3.1% to -0.9%). (2) When adding noises on 800 capnograms, VD(aw) was underestimated whenever the FM was used thereby creating a bias between -5.54 and -1.28 ml at standard deviations (SD) of 0.1-1.8 ml (P < 0.0001). FA-LMA calculations of VD(aw) were close to the simulated values with the bias ranging from -0.21 to 0.16 ml at SD from 0.1 to 0.4 ml. The FM overestimated S(III) and showed more bias (0.0041-0.0078 mmHg/ml, P < 0.0001) than the FA-LMA (0.0002-0.0030 mmHg/ml). When calculating VD(aw) from patients, variability was less with the FA-LMA leading to mean variation coefficients of 0.0102, 0.0111 and 0.0123 compared to the FM (0.0243, 0.0247 and 0.0262, P < 0.001) for 0, 5 and 10 cmH(2)O of PEEP, respectively. The FA-LMA also showed less variability in S(III) with mean variation coefficients of 0.0739, 0.0662 and 0.0730 compared to the FM (0.1379, 0.1208 and 0.1246, P < 0.001) for 0, 5 and 10 cmH(2)O of PEEP, respectively. CONCLUSIONS: The Functional Approximation based on a Levenberg-Marquardt Algorithm showed less bias and dispersion compared to the traditional Fowler's method when calculating VD(aw) and S(III).},\n bibtype = {article},\n author = {Tusman, Gerardo and Scandurra, Adriana and Böhm, Stephan H and Suarez-Sipmann, Fernando and Clara, Fernando},\n journal = {Journal of clinical monitoring and computing},\n number = {4}\n}","author_short":["Tusman, G.","Scandurra, A.","Böhm, S., H.","Suarez-Sipmann, F.","Clara, F."],"urls":{"Paper":"http://bibbase.org/service/mendeley/6d353feb-efe4-367e-84a2-0815eb9ca878/file/705dccee-5653-4ea6-42cd-7a69105f1c7f/Model fitting of volumetric capnograms improves calculations of airway dead space and slope of phase III..pdf"},"bibbaseid":"tusman-scandurra-bhm-suarezsipmann-clara-modelfittingofvolumetriccapnogramsimprovescalculationsofairwaydeadspaceandslopeofphaseiii-2009","role":"author","keyword":["abstract","background","co 2","dead space","from the 1 department","hospital privado de","lung recruit-","ment","monitoring","of anesthesiology","peep","phase iii slope","volumetric capnography"],"downloads":0},"bibtype":"article","biburl":null,"creationDate":"2014-10-29T08:42:17.907Z","downloads":0,"keywords":["abstract","background","co 2","dead space","from the 1 department","hospital privado de","lung recruit-","ment","monitoring","of anesthesiology","peep","phase iii slope","volumetric capnography"],"search_terms":["model","fitting","volumetric","capnograms","improves","calculations","airway","dead","space","slope","phase","iii","tusman","scandurra","böhm","suarez-sipmann","clara"],"title":"Model fitting of volumetric capnograms improves calculations of airway dead space and slope of phase III.","year":2009}